• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH VIPER PRIME CFXFN XTB 7X45MM S; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH VIPER PRIME CFXFN XTB 7X45MM S; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD Back to Search Results
Model Number 186770445S
Medical Device Problem Codes Break (1069); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/08/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional procodes: kwp, kwq.Complainant part is not expected to be returned for manufacturer review/investigation.Customer quality investigation: the complaint device was not received for investigation.The following investigation is based on the image provided.The image was reviewed, and the complaint condition could be confirmed since the screw appears to be stuck on the end of the unknown screwdriver.The screw also appears to be missing the tulip head and tabs.Due to how the device is assembled, the tulip head and tabs cannot separate from the shank without breakage in the screw.Since the device was not returned, a dimensional inspection and a functional test were not able to be performed.A manufacturing record evaluation could not be performed as the lot number could not be determined from the image provided.A definitive assignable root cause could not be determined based on the provided information.During the investigation, no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from (b)(6) reports an event as follows: it was reported that during an mis procedure with viper prime on (b)(6) 2019, a screw became stuck in the screwdriver and could not be removed.The stylet was also stuck.The surgeon needed to use other devices.Procedure was completed successfully with a delay of twenty (20) minutes.There were no patient consequences.This report is for a viper prime screw.This is report 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER PRIME CFXFN XTB 7X45MM S
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key9615942
Report Number1526439-2020-00420
Device Sequence Number585887
Product Code NKB
UDI-Device Identifier10705034530253
UDI-Public(01)10705034530253
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K173095
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number186770445S
Device Catalogue Number186770445S
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/09/2020
Supplement Date Received by Manufacturer01/09/2020
Initial Report FDA Received Date01/22/2020
Supplement Report FDA Received Date01/22/2020
Patient Sequence Number1
Patient Age72 YR
-
-