Catalog Number H965457591
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 01/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was reported that the disposable device is not available to be returned to the manufacturer for evaluation.An investigation into the root cause of this incident is currently in progress.The results of the investigation will be sent via a follow up medwatch.Complaint # (b)(4).
 
Event or Problem Description
Received user medwatch from angiodynamics' customer on january 10 2020, reference number (b)(4).Per the report event description: fracture of mini stick max wire during case, had to be snared and retrieved.It was reported the defective disposable device is not available for return to the manufacturer.
 
Additional Manufacturer Narrative
As the reported defective device was not returned, angiodynamics is unable to perform a device evaluation.The supplier of this device was notified of the event via (b)(4) and requested to provide a device history record review.Per the supplier response: there was no observation made during records review that identify any contributing factors to the failure reported.All inspections results were recorded within specification limits.The product specifications were met with no non-conformities reported for manufacturing and inspection.A general training review was performed.It was verified that all personal was properly trained in all manufacturing process steps and in quality inspection as well.A review of the device packaging history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.Labeling review: direction for use {dfu}: warning.Contents supplied sterile using an ethylene oxide (eo) process.Do not use if sterile barrier is damaged.If damage is found, call your navilyst medical representative.Inspect prior to use to verify that no damage has occurred in shipping.For single patient use only.Do not reuse, reprocess or resterilize.Reuse, reprocessing or resterilization may compromise the structural integrity of the device and/or lead to device failure which, in turn, may result in patient injury, illness or death.Operational instructions.1.Prior to insertion, flush all components with saline or heparinized/saline.2.Gain percutaneous access with the 21 g (0.9 mm) vascular introducer needle.3.Advance the 0.018 inch (0.46 mm) guidewire through the needle.Caution: the guidewire should not be withdrawn through the introducer needle.If the guidewire must be withdrawn while still inside the needle, simultaneously remove both the needle and the wire as a unit to prevent the needle from damaging the guidewire.4.Withdraw the introducer needle, leaving the 0.018 inch (0.46 mm) guidewire in place.5.Advance the coaxial assembly over the 0.018 inch (0.46 mm) guidewire.6.Simultaneously remove the dilator and 0.018 inch (0.46 mm) guidewire, while holding the sheath portion of the assembly in place.7.Advance a 0.035 inch (0.89 mm) or 0.038 inch (0.97 mm) guidewire into the sheath.8.Remove the sheath, leaving the 0.035 inch (0.89 mm) or 0.038 inch (0.97 mm) wire.A review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINI STICK MAX
Common Device Name
VESSEL DILATOR
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny NY 12801
MDR Report Key9617317
Report Number1317056-2020-00003
Device Sequence Number11447933
Product Code DRE
UDI-Device IdentifierH965457591
UDI-PublicH965457591
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K132713
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 03/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/30/2022
Device Catalogue NumberH965457591
Device Lot Number5530160
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/10/2020
Supplement Date Received by Manufacturer01/10/2020
Initial Report FDA Received Date01/22/2020
Supplement Report FDA Received Date03/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient Weight91
-
-