It was reported during use while in monitoring mode, the unit alarmed for "mechanical ventilation not available." based on the preliminary logfile analysis, it was found that the device was not used in monitoring mode but in pressure mode when the ventilator failed.There was no patient injury reported.
|
Based on the available log file, the reported case could be reconstructed.It was found that during the case in question, a sporadic failure of the inspiratory pressure sensor of the vgc (ventilation and gas controller) was detected by the device.In consequence, the ventilator initiated an autonomous shutdown while changing mode (pressure mode) to man/spont accompanied by the corresponding "ventilator fail" alarm as specified.In this case, manual ventilation as well as monitoring function remains unaffected.After the case in question, the device was rebooted and passed the power-on self-test.As further reported, the unit remained in use in follow-up to the event with no further issues reported.Dräger finally concludes that the device reacted as specified upon a sporadic failure of the inspiratory pressure sensor.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|