According to the reporter, during a right colon hemicolectomy procedure, there was an incomplete staple line on the central area and poor staple formation using the first reload and handle for tissue anastomosis.Another reload was used to repair the staple line but the device could not be fired.It was reported that the surgeon was able to press the green button but was unable to squeeze the handle.A third reload was used together with a new handle to complete the procedure.There was no patient injury.
|
|
H6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was returned for evaluation.Inspection and testing found that the firing knobs were retracted, and the articulation lever was in neutral position.Visual inspection and testing found no potentially contributing factors.A device-related casual link was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are also thoroughly reviewed prior to release to ensure that products meet all medtronic quality specifications.Improvements have been made to mitigate the issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|