• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. SHOULDER PRIMARY HUMERAL STEM,; HEAD, HUMERAL, OFFSET, 46-16

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. SHOULDER PRIMARY HUMERAL STEM,; HEAD, HUMERAL, OFFSET, 46-16 Back to Search Results
Model Number 520-46-116
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/24/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Revision surgery - due to dislocation.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was reported as dislocation.The previous surgery and the surgery detailed in this event occurred 14 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) shows that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to dislocation.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.Due to short time between the previous and revision surgery, it seems that the event may possibly occurred due to patient noncompliance with medical instruction or improper implant selection.There are multiple factors that may contribute to an event that are outside the control of djo surgical such as poor bone density, patient bone deterioration, inadequate soft tissue support, excessive range of motion, prolonged overhead activities, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOULDER PRIMARY HUMERAL STEM,
Common Device Name
HEAD, HUMERAL, OFFSET, 46-16
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445 TX 78758 5445
MDR Report Key9623117
Report Number1644408-2020-00003
Device Sequence Number13578896
Product Code KWS
UDI-Device Identifier00888912025034
UDI-Public(01)00888912025034
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K080402
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/09/2024
Device Model Number520-46-116
Device Catalogue Number520-46-116
Device Lot Number923C1135
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/24/2019
Supplement Date Received by Manufacturer01/30/2020
Initial Report FDA Received Date01/23/2020
Supplement Report FDA Received Date02/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-00-000, LOT 878C1775; 520-01-008, LOT 454G1111; 520-00-000, LOT 878C1775; 520-01-008, LOT 454G1111
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
-
-