It was reported that this patient with heart block was implanted with a pacemaker and implant testing produced stable measurements and there were no issues with the procedure.Additionally, device testing the following day was unremarkable for any issues.However, data from a telemetry strip occurring after the last device check showed pacing inhibition resulting in pacing below the lower rate limit.A boston scientific technical services consultant reviewed the data and discussed possible explanations for the clinical observations.No adverse patient effects were reported.
|