| Model Number |
MMT-723LNAL |
| Medical Device Problem Codes |
Failure to Cycle (1142); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had a compromised force sensor system alarm.The customer stated during seating the force sensor did not detect the reservoir.The device was not bumped or dropped prior to the alarm and the drive support cap appeared flush.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test but a33 alarm during the basic occlusion test due to protruded drive support disk.Unable to perform the occlusion, prime/a33 and excessive no delivery test due to a33 alarm.
|
| |
|
Search Alerts/Recalls
|