Model Number 77722
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 01/13/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation summary: visual inspection shows evidence of damage/hole in the body of the device.During the cleaning process there was a leak observed from the damage/hole.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that during use of an eopa arterial cannula, it was noted that there was a pinhole leak from the cannula throughout the case.The customer put a towel over the cannula to complete the case.The product was used to complete the case.Patient lost 100ml of blood a transfusion was not required. .
 
Additional Manufacturer Narrative
Conclusion: the product was returned from the field damaged; the quality checks that are in place would not allow for this type of defect to pass through our system and likely occurred after it left our facility.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EOPA ARTERIAL CANNULA
Common Device Name
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
MDR Report Key9626747
Report Number2184009-2020-00005
Device Sequence Number10857479
Product Code DWF
UDI-Device Identifier00643169340480
UDI-Public00643169340480
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K031037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Model Number77722
Device Catalogue Number77722
Device Lot Number2018010843
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2020
Initial Date Received by Manufacturer 01/13/2020
Supplement Date Received by Manufacturer02/19/2020
Initial Report FDA Received Date01/24/2020
Supplement Report FDA Received Date02/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Weight91
-
-