Additional Manufacturer Narrative
|
|
Bayer product analysis received and examined the returned product.Visual examination confirmed the presence of a foreign material within the syringe barrel.Our investigation determined that the foreign material was most likely introduced during the manufacturing process at the third-party supplier and went undetected.A 12-month review of complaint history determined the complaint rate to be 0.49 complaints per million (cpm).
|
| |
|
Event or Problem Description
|
|
The customer reported the following: the customer observed a foreign material in the saline syringe barrel of a ssqk 65/115vs syringe kit.No injury or adverse event was reported.
|
| |
|