• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EBI, LLC. ELECTRODES, LT-4500

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EBI, LLC. ELECTRODES, LT-4500 Back to Search Results
Model Number N/A
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).Date of event: the event occurred sometime in (b)(6) 2020.The customer has indicated that the product will not be returned to zimmer biomet for investigation because the patient is still using the product.The investigation is in process.Once the investigation has been completed, a follow up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0002242816-2020-00013 and 0002242816-2020-00014.The device has not been returned.
 
Event or Problem Description
It was reported that the patient was experiencing pain from the spinalpak stimulator.The patient was using the spinalpak stimulator with the lt-4500 and 72r electrodes.The patient stated that the pain was on the right side of her back in the kidney area and across the hip.The patient stated that the pain is below the skin and the pain level is a 10, on a scale of 1-10.The patient stopped using the stimulator and the pain did not go away.The patient called their doctor and was prescribed ibuprofen and advised not to use the stimulator for a week.The patient believes that the pain was from the inflammation of the incision.No additional patient consequences have been reported.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The device was not returned to zimmer biomet for evaluation.The reported event was unable to be confirmed due to limited information received from the customer.The device history record was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.Device evaluated by manufacturer updated to no.Method code updated to 3331: method code updated to 4114: device not returned.Results code updated to 3221: no findings available.Conclusions code updated to 4315: cause not established.
 
Event or Problem Description
It was reported that the patient was experiencing pain from the spinalpak stimulator.The patient was using the spinalpak stimulator with the lt-4500 and 72r electrodes.The patient stated that the pain was on the right side of her back in the kidney area and across the hip.The patient stated that the pain is below the skin and the pain level is a 10, on a scale of 1-10.The patient stopped using the stimulator and the pain did not go away.The patient called their doctor and was prescribed ibuprofen and advised not to use the stimulator for a week.The patient believes that the pain was from the inflammation of the incision.It was reported that no further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTRODES, LT-4500
Common Device Name
ELECTRODES
Manufacturer (Section D)
EBI, LLC.
399 jefferson road
parsippany NJ 07054
MDR Report Key9627090
Report Number0002242816-2020-00012
Device Sequence Number3168864
Product Code LOF
UDI-Device Identifier00812301020188
UDI-Public00812301020188
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P850022/S017
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number106130-12
Device Lot Number913014
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/08/2020
Supplement Date Received by Manufacturer05/27/2020
Initial Report FDA Received Date01/24/2020
Supplement Report FDA Received Date06/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Weight98
-
-