• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD CARELINK SMARTSYNC BASE; ANALYZER, PACEMAKER GENERATOR FUNCTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD CARELINK SMARTSYNC BASE; ANALYZER, PACEMAKER GENERATOR FUNCTION Back to Search Results
Model Number 24970A
Device Problems Computer Software Problem (1112); Communication or Transmission Problem (2896); Computer Operating System Problem (2898)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/08/2019
Event Type  malfunction  
Manufacturer Narrative
Product analysis : analysis confirmed original complaint.The behavior is linked to a known issue, rpi 108712.The interrogation hangs when initiated while the application is in background.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during use of an electrocautery device, the programmer had been placed on the cart next to the cautery machine.When the physician started using cautery, the programmer displayed a dotted line for the connection status.When cautery was turned off, the connection resumed.When the physician continued use of cautery, the application completely crashed and the home screen was displayed.In addition, when first pulling up the programmer, and trying to pair the patient connector, the application completely froze.A spinning hour glass displayed, and the application was frozen for five minutes.The user closed the application to restart.The programmer remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK SMARTSYNC BASE
Type of Device
ANALYZER, PACEMAKER GENERATOR FUNCTION
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9629918
MDR Text Key176381295
Report Number3004593495-2020-00061
Device Sequence Number1
Product Code DTC
UDI-Device Identifier00643169957657
UDI-Public00643169957657
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number24970A
Device Catalogue Number24970A
Was Device Available for Evaluation? No
Date Manufacturer Received01/21/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/22/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-