| Model Number |
MMT-723NAP |
| Medical Device Problem Codes |
No Display/Image (1183); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.(b)(4).
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that insulin pump was not able to turn on.Customer dropped the insulin pump.Customer reported no visible pump damage.No harm requiring medical intervention was reported.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with blank display due to moisture damage at electronic assembly.Also, moisture damage motor during visual inspection.Unable to perform operating currents, self test, rewind test and displacement test due to blank display.Device received with cracked belt clip slot and cracked reservoir tube lip.
|
| |
|