| Model Number |
N/A |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the instrument was missing a bearing.This was found during an inventory check.No patient involvement.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Verified item lot combination and reviewed manufacturing date as tasp lot 62117747 exhibits signs of repeated use (nicked or gouged) and has one of components 1 and 2 disassembled / missing.The missing component was not returned.The dhr was reviewed and no discrepancies relevant to the reported event were found.The device is confirmed to have been a part of a previous investigation.Actions were initiated to recommend against using ultrasonic cleaning as a sterilizing technique for these devices.These devices were subsequently re-designed to account for this failure mode with a new locking mechanism.It was determined that these devices were manufactured prior to this design change.The root cause is considered to be a previously addressed design issue.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|