• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065751155
Device Problem Suction Problem (2170)
Patient Problem Burn, Thermal (2530)
Event Date 01/15/2020
Event Type  Injury  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A healthcare professional reported that there was no aspiration on first pass during sculpt and wound burn was observed.Sutures were placed and a new incision was made.The handpiece, tip and cassette were replaced and the procedure was completed.
 
Manufacturer Narrative
Additional information provided in d.10, h.3, h.6 and h.10.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The wet returned sample was visually inspected and no obvious defects were found that would have contributed to the reported event.A console representing the current software version was used to test the sample.The ball in the drip chamber¿s check valve moved freely per specification.The sample could prime and tune with the ultrasonic handpiece successfully.The irrigation and aspiration pressure were measured at multiple set points and met specifications.No message code appeared on the screen.Fluid flowed from balanced salt solution (bss) to the drain bag without any interfering.No leakage was detected from the pump elastomer or on the pump area of the fluidics module.Cleaning process was able to be performed after functional test was completed.The sample passed functionality.The root cause of the customer's complaint could not be established; the returned sample met specifications.After investigation of this complaint, it has been determined that this sample met specifications; therefore, no corrective action is required at this time.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: 2020-03871.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key9632730
MDR Text Key176435782
Report Number1644019-2020-00055
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2021
Device Catalogue Number8065751155
Device Lot Number2290941H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2020
Date Manufacturer Received04/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CONSTELLATION VISION SYSTEM; CONSTELLATION VISION SYSTEM
Patient Outcome(s) Required Intervention;
-
-