| Model Number |
10607 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
12/29/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.
|
| |
|
Event or Problem Description
|
|
It was reported that stent fracture occurred.The target lesion was located in the right coronary artery.A 3.50x38mm synergy ii drug-eluting stent (des) was deployed for treatment.However, during an intravascular ultrasound, a fracture was noted.The lesion was ballooned and a 35x12mm synergy ii des was successfully deployed.There were no patient complications nor injuries reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device is a combination product.Device evaluated by mfr.: the device was not returned to the cis for analysis but a single still angiographic image was returned attached to complaint record.The image was reviewed.The image appeared to be a photograph taken of the monitor displaying an angiographic image.The image showed a guide catheter positioned at the right coronary artery (rca) ostium, with contrast flow through the rca.A pacing wire was evident in the heart.A flow narrowing/constriction was noted in the proximal rca.It could not be confirmed in the image if a stent was present or if it was fractured or not.The single angiographic image provided could not be used to confirm or refute the allegation of stent fracture.H3 other text : angiographic image.
|
| |
|
Event or Problem Description
|
|
It was reported that stent fracture occurred.The target lesion was located in the right coronary artery.A 3.50x38mm synergy ii drug-eluting stent (des) was deployed for treatment.However, during an intravascular ultrasound, a fracture was noted.The lesion was ballooned and a 35x12mm synergy ii des was successfully deployed.There were no patient complications nor injuries reported.
|
| |
|
Search Alerts/Recalls
|