Model Number 365712
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the catheter fell out of the patient with a ruptured balloon.The catheter was replaced with the same foley catheter and all of the balloon pieces were accounted for.It was later reported via email on 15 january 2020 from the complainant, the balloon was intact with no missing pieces.
 
Additional Manufacturer Narrative
The reported event was confirmed, however, the cause is unknown.Evaluation observed that the balloon had burst as irregular with no pieces of sac missing.The sample was evaluated under a microscope and no conditions were found that could be associated with the reported event.However, the exact cause on how and when problem occurred could not be determined.A potential root cause for this failure mode could be user related (example: contact with sharp object)/ / exposure to petrolatum based products mechanical failure/operator error.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "bard® foley catheter caution: this product contains natural rubber latex which may cause allergic reactions.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Sterile unless package is opened or damaged.Do not use if package is opened or damaged.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage latex and may cause balloon to burst.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Single patient use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer syringe.Do not use needle.Visually inspect the product for any imperfections or surface deterioration prior to use.Recommended inflation capacities 3cc balloon: use 5cc sterile water 5cc balloon: use 10cc sterile water 15cc balloon: use 20cc sterile water 20cc balloon: use 25cc sterile water 30cc balloon: use 35cc sterile water 40cc balloon: use 45cc sterile water 75cc balloon: use 80cc sterile water do not exceed recommended capacities note: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulations.Bard is a registered trademark of c.R.Bard, inc.Or an affiliate.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact an adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the catheter fell out of the patient with a ruptured balloon.The catheter was replaced with the same foley catheter and all of the balloon pieces were accounted for.It was later reported via email on 15 january 2020 from the complainant, the balloon was intact with no missing pieces.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® LUBRICATH® FOLEY CATHETER
Common Device Name
FOLEY CATHETHER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9633605
Report Number1018233-2020-00541
Device Sequence Number14241492
Product Code EZC
UDI-Device Identifier00801741077722
UDI-Public(01)00801741077722
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K910846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/28/2023
Device Model Number365712
Device Catalogue Number365712
Device Lot Number8RA008
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2020
Initial Date Received by Manufacturer 01/10/2020
Supplement Date Received by Manufacturer02/21/2020
Initial Report FDA Received Date01/27/2020
Supplement Report FDA Received Date02/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-