| Model Number |
FFRCT 2.47.0.1 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of heartflow's quality monitoring process, we identified a potential false negative.(b)(4): the investigation identified a potential false negative in the rca region; after the initial analysis was delivered, a second analysis completed as part of ongoing quality review resulted in a decrease in the ffrct value from 0.88 to 0.67 in the rca region.Heartflow was able to confirm with the ordering physician that the reanalysis did not result in consequences or impact to the patient.
|
| |
|
Event or Problem Description
|
|
Heartflow identified a potential false negative result.
|
| |
|