| Model Number |
37370DL |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Patient Problem/Medical Problem (2688)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The evaluation found the tip of the electrode is broken, the application tip has not been sent in.The shaft is bent at the end.Furthermore, the shaft has several scratches and dents.Presumably the electrode was bent and bent back.No indication of a material or manufacturing defect found during investigation.The root cause is most likely overloading of the instrument due to excessive force.
|
| |
|
Event or Problem Description
|
|
Per the factory in (b)(4), allegedly there was an event which occurred in (b)(6) that the hook broke off and was intraoperative.
|
| |
|