COOK INC GUNTHER TULIP FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
Catalog Number IGTCFS-65-FEM |
Device Problems
Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Internal Organ Perforation (1987); Pain (1994); Perforation of Vessels (2135); Anxiety (2328); Distress (2329); No Information (3190)
|
Event Type
Injury
|
Manufacturer Narrative
|
Occupation: non-healthcare professional.Investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.Catalog number and lot number are unknown, however, the alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.
|
|
Event Description
|
It is alleged that the patient received a gunther tulip on (b)(6) 2009.It is alleged that the patient was injured without further explanation.Hospital and medical records have been requested but not yet provided.
|
|
Event Description
|
Per a (b)(6) 2019 ct (computed tomography) abdomen: "findings: there is an ivc filter.The filter has a slight tilt of 11 degrees to the patient's left.The tip of the filter is 2 millimeters below the lowest renal vein.There is no evidence for filter migration.There is perforation of 4 struts of the filter into the surrounding mesentery.There is no aortic perforation, small bowel perforation, adrenal perforation, pancreatic perforation, renal vein or kidney perforation, lobar spine or psoas muscle perforation or perforation of the diaphragm.There is no filter struck fracture or bending.There is no vena cava stenosis.The apex of the filter is not touching the ivc wall".
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional information: a2, b1, b2, b5, b6, h6 (patient and device codes) h6 (device code): appropriate term/code not available (3191) for alleged device perforation.H6 (device code): appropriate term/code not available (3191) for alleged device tilt investigation: investigation is reopened due to additional information provided.The following allegations have been investigated: vena cava (vc)/organ perforation, tilt the reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.A total of 20 devices were manufactured in the reported lot.To date, one other complaint has been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to current controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Patient allegedly received an implant on (b)(6) 2009 due to deep vein thrombosis (dvt).Patient is alleging device tilt and vena cava perforation.Patient notes and further alleges experiencing "pain, suffering, loss of enjoyment of life, distress", and anxiety.
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Unknown if the reported pain, suffering, loss of enjoyment of life, distress, and anxiety are directly related to the filter and unable to identify a corresponding failure mode at this point in time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Search Alerts/Recalls
|
|
|