| Model Number |
226820 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
01/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was evaluated by the manufacturer, karl storz in (b)(4).As per the evaluation: the round knife is broken off distally and the broken part is missing.Corrosion can be seen on the fracture surface, which indicates pre-damage.Excessive force, such as leveraging, can lead to damage.The knife must be checked tor damage before each use and should have been detected.Furthermore, not too much force should be applied to the knife.
|
| |
|
Event or Problem Description
|
|
Per a vigilance report received from the factory in (b)(4), allegedly, there was an event that occurred at a healthcare facility in (b)(6).The distal end of the instrument, a round knife, broke during the procedure and the procedure was aborted.Customer confirmed that the broken fragment did not remain in the patient.
|
| |
|