STRYKER INSTRUMENTS-KALAMAZOO INTERPULSE HANDPIECE SET WITH SUCTION TUBING; LAVAGE, JET Back to Search Results
Catalog Number 0210100000
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Date of Event 12/31/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events: 8 events were reported for this quarter.Product return status: 7 devices were received.1 device was not available for evaluation.Event confirmation status: 7 reported events were confirmed.Evaluation results: 7 devices were found to be affected by a broken battery port.Additional information: 8 devices were labeled for single-use.8 devices were not reprocessed or reused.
 
Event or Problem Description
This report summarizes 8 malfunction events in which the device was reportedly leaking.8 events had patient involvement; no patient impact.
 
Additional Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Supplemental rationale corrected data: b5, h10 8 events were originally reported for this failure mode during the reporting quarter.- 1 event was inadvertently excluded.- 9 reported events are included in this follow-up record.Product return status 8 devices were received.1 device was not available for evaluation.Event confirmation status 8 reported events were confirmed.Evaluation results 8 devices were found to be affected by a broken battery housing port/stem.
 
Event or Problem Description
This report summarizes 9 malfunction events in which the device was reportedly leaking.9 events had patient involvement; no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERPULSE HANDPIECE SET WITH SUCTION TUBING
Common Device Name
LAVAGE, JET
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key9636407
Report Number0001811755-2020-00146
Device Sequence Number11460924
Product Code FQH
UDI-Device Identifier34546540144219
UDI-Public34546540144219
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized9
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number0210100000
Device Lot NumberVMSR
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/31/2019
Supplement Date Received by Manufacturer12/31/2019
Initial Report FDA Received Date01/28/2020
Supplement Report FDA Received Date04/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-