Model Number 7208008000
Medical Device Problem Code Fail-Safe Problem (2936)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 12/31/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events: 3 events were reported for this quarter.Product return status: 3 devices were received.Event confirmation status: 2 reported events were confirmed.1 reported event was not confirmed.Evaluation results: 3 devices were found to be affected by corrosion.Additional information: 3 devices were not labeled for single-use.3 devices were not reprocessed or reused.
 
Event or Problem Description
This report summarizes 3 malfunction events in which the device experienced a fail-safe problem that could result in unintended activation.3 events had no patient involvement; no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYSTEM 7 V-NOTCH SAGITTAL SAW
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key9636649
Report Number0001811755-2020-00280
Device Sequence Number13541597
Product Code HWE
UDI-Device Identifier07613252251040
UDI-Public07613252251040
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized3
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number7208008000
Device Catalogue Number7208008000
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/31/2019
Initial Report FDA Received Date01/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-