STRYKER INSTRUMENTS-KALAMAZOO CORDLESS DRIVER 4 HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Model Number 4405000000
Device Problem Fluid/Blood Leak (1250)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 12/31/2019
Event Type  malfunction  
Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events: 18 events were reported for this quarter.Product return status: 18 devices were received.Event confirmation status: 13 reported events were confirmed.5 reported events were not confirmed.Evaluation results: 2 devices were found to be affected by a worn or damaged handle assembly.1 device was found to be affected by a worn or damaged handle assembly and corrosion.5 devices were found to be affected by non-stryker components.6 devices did not have a root cause established.4 devices had no problem found.Additional information: 18 devices were not labeled for single-use.18 devices were not reprocessed or reused.
 
Event Description
This report summarizes 18 malfunction events in which the device was reportedly leaking.17 events had no patient involvement; no patient impact.1 event had no known patient involvement or patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORDLESS DRIVER 4 HANDPIECE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
zach baker
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key9637270
MDR Text Key177623070
Report Number0001811755-2020-00319
Device Sequence Number1
Product Code HWE
UDI-Device Identifier07613252481355
UDI-Public07613252481355
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported18
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number4405000000
Device Catalogue Number4405000000
Was Device Available for Evaluation? No
Date Manufacturer Received12/31/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-