• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 4.75 HEALIX ADVANCE KNTLSS PK; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 4.75 HEALIX ADVANCE KNTLSS PK; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE Back to Search Results
Model Number 222333
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 01/06/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).The expiration date is unknown.
 
Event or Problem Description
It was reported by affiliate via complaint submission tool that during an unknown procedure, while using a 4.75mm healix advance knotless peek anchor, when the surgeon was going to implant the anchor it popped back out and it would not stay in the bone.There was a surgical delay of 7 minutes and procedure was completed using a new implant.No patient consequences reported.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: correction: a1: patient identifier was inadvertently missed on the initial report.The patient identifier is (b)(6).Additional information: h6: method codes: device history lot
=
> a manufacturing record evaluation was performed for the finished device lot number (6l19156), and no non-conformances were identified.Investigation summary
=
> the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.A manufacturing record evaluation was performed for the finished device lot number (6l19156), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.75 HEALIX ADVANCE KNTLSS PK
Common Device Name
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9637945
Report Number1221934-2020-00336
Device Sequence Number14181410
Product Code MBI
UDI-Device Identifier10886705023455
UDI-Public10886705023455
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K130539
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number222333
Device Catalogue Number222333
Device Lot Number6L19156
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/16/2020
Supplement Date Received by Manufacturer02/13/2020
Initial Report FDA Received Date01/28/2020
Supplement Report FDA Received Date02/14/2020
Patient Sequence Number1
-
-