| Model Number |
D001 |
| Medical Device Problem Codes |
High impedance (1291); Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/06/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As no further information regarding this event is expected, our investigation is complete.This report will be updated should additional information be provided.
|
| |
|
Event or Problem Description
|
|
It was reported that this patient's implanted system had delivered a shock which resulted in code 1005 being exhibited which is indicative of a shock into an open lead.A high out of range shock impedance measurement of greater than 125 ohms was also observed.The system has been reprogrammed and remains in service.No adverse patient effects were reported.
|
| |
|