| Model Number |
VBJR081502A |
| Medical Device Problem Code |
Activation Problem (4042)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Review of the manufacturing records indicated the device met pre-release specifications.No device or images were provided to gore for evaluation.Therefore, based on the engineering evaluation performed, no root cause was identified.
|
| |
|
Event or Problem Description
|
|
The following was reported to gore: on (b)(6) 2020 the patient presented with thrombosis in the brachial artery.An 8mm x 15cm gore® viabahn® endoprosthesis was to be deployed as a brachial artery to brachial vein fistula.However, during deployment, the device started to "bowstring" and would not deploy after approximately 1cm of the device was expanded.The device was removed and an 8mm x 15cm gore® viabahn® endoprosthesis was successfully deployed to complete the procedure.The patient tolerated the procedure and is doing well.
|
| |
|