Model Number TMICL12.1
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted a 12.1mm tmicl12.1 implantable collamer lens, (-7.5/+1.5/090 (sphere/cylinder/axis), into the patient's right (od) eye on (b)(6) 2019.Reportedly, approximately 1-month postop the patient has more cylinder than she started with.It was determined that the icl is sitting vertically in the eye.On (b)(6) 2020 the lens was repositioned.1-week post-reposition, the uncorrected vision is 20/40.
 
Additional Manufacturer Narrative
H6-method code information should have been included in initial medwatch report: a lens work order search was performed.No similar complaint type events reported for units within the same lot.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key9639455
Report Number2023826-2020-00187
Device Sequence Number319687
Product Code QCB
UDI-Device Identifier00841542104333
UDI-Public00841542104333
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2021
Device Model NumberTMICL12.1
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/06/2020
Supplement Date Received by Manufacturer02/10/2020
Initial Report FDA Received Date01/28/2020
Supplement Report FDA Received Date02/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-