MENTOR TEXAS MENTOR SILTEX CONTOUR PROFILE MODERATE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Model Number |
3542913 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
12/01/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: deflation.Concomitant medical products: (right) 350cc mentor siltex contour profile moderate saline catalog: 3542913 lot: 5858159 sn: (b)(4).Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) caucasian female patient who underwent breast augmentation revision with two 350cc mentor siltex contour profile moderate saline breast prostheses, suffered left breast deflation, post-operatively.As a result, the patient underwent implant explantation on (b)(6) 2020.
|
| |
|
Additional Manufacturer Narrative
|
|
On 2/6/2020, the mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2020, the product investigation was completed.A manufacturing record evaluation (mre) was performed for the complaint device.The manufacturing record evaluation (mre) was reviewed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Device investigation summary: during visual inspection of the device no apparent damage could be observed.Leak testing revealed there was leakage from the valve.The analysis was¿unable to determine the root cause for the leaking valve.Excessive manipulation may have caused the valve leak.The ifu states ¿not to manipulate (i.E., squeeze) the valve excessively, which may cause valve leakage¿.However, this cannot be conclusively determined.The manufacturing records were reviewed, and the manufacturing criteria were met prior to the release of this lot.The second product received (product code 3542913 and lot number 5858159) is a concomitant device, therefore no further analysis is required.It should be noted that as part of our quality process, each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Manufacturer¿s reference number: (b)(4).
|
| |
|