| Model Number |
SR-200-NPS |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Patient Problem/Medical Problem (2688); Vascular Dissection (3160)
|
| Date of Event |
09/16/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product in complaint was not returned to the manufacturer for analysis.A review of the manufacturing records for this device was completed and no issues were identified that could have lead to the adverse event reported.Complaints will continue to be monitored for any trends.On 17dec2019 silk road medical contacted osb to inform them of medical device reports that were not reported within the mandatory reporting timeframe.It is noted here, per fda's request, that this mdr was identified as part of a complaint handling remediation effort associated with capa-(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient enrolled in a clinical study underwent a transcarotid revascularization procedure on (b)(6) 2016.The patient experienced a dissection of the common carotid artery (cca) backwall during introduction of the neuroprotection system.The dissection was successfully treated with the placement of a second carotid stent.No additional details were provided.
|
| |
|