Model Number TSH
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/24/2019
Event Type  malfunction  
Manufacturer Narrative
The investigation did not identify a product problem. The cause of the event could not be determined. Results with the different analyzers from different vendors can generate different values. This relates to the overall setups of the assays, the antibodies used and, differences in reference materials/methods and the standardization methodology used. From the information provided, a general reagent issue can most likely be excluded. This event occurred in (b)(6). (b)(4).
 
Event Description
The initial reporter received questionable elecsys tsh assay results for one patient from cobas e 801 module serial number (b)(4). The sample was submitted for investigation and was tested on architect analyzer and a cobas e 801 module. Refer to the attachment to the medwatch for all patient data. This medwatch is for tsh. Refer to the medwatch with patient identifier (b)(4) for the ft3 assay and (b)(4) for the ft4 assay. The customer reported out the results to a physician.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameELECSYS TSH ASSAY
Type of DeviceRADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM 68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key9641396
MDR Text Key219782345
Report Number1823260-2020-00227
Device Sequence Number1
Product Code JLW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K961491
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/29/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberTSH
Device Catalogue Number07028091190
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/14/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-