DEPUY MITEK LLC US VAPR 3.5MM SIDE STR -EA; ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE USE Back to Search Results
Model Number 225301
Medical Device Problem Code Blocked Connection (2888)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/09/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).Udi: (b)(4).
 
Event or Problem Description
It was reported by the affiliate that during an unknown procedure treating the foot joint, when the vapr 3.5 mm side str ea was connected to the generator it was not recognized.The procedure was completed using a replacement.No patient consequence and no surgical delay was reported.No additional information was provided.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary
=
> according to the information provided, it was reported that the electrode was not recognized.The complaint device is not being returned, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.Since the complaint device was not returned, we cannot determine a root cause for the reported failure.If the device is received in the future, we will reopen the complaint and perform the investigation as appropriate.A manufacturing record evaluation was performed for the finished device lot number [u1902051], and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot
=
> a manufacturing record evaluation was performed for the finished device lot/serial number (b)(6), and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAPR 3.5MM SIDE STR -EA
Common Device Name
ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE USE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9641485
Report Number1221934-2020-00350
Device Sequence Number2139406
Product Code GEI
UDI-Device Identifier10886705009251
UDI-Public10886705009251
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K082643
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2024
Device Model Number225301
Device Catalogue Number225301
Device Lot NumberU1902051
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/16/2020
Supplement Date Received by Manufacturer02/21/2020
Initial Report FDA Received Date01/29/2020
Supplement Report FDA Received Date02/24/2020
Patient Sequence Number1
-
-