| Model Number |
7005 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/16/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The balloon was tightly folded.There were numerous hypotube kinks.The hypotube was separated 19.5cm from the hub.The separated ends of the hypotube was ovaled indicating the hypotube was kinked prior to separating.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 10jan2020.It was reported that shaft kink occurred.The 93% stenosed, 4x10mm target lesion was located in moderately tortuous and moderately calcified left anterior descending artery.A 4.0mm x 8mm quantum maverick balloon catheter was advanced for dilatation.However, during procedure the balloon delivery shaft was kinked.The procedure was completed with different device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed shaft detached/separation.
|
| |
|