| Model Number |
7010 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Returned product consisted of a quantum maverick balloon catheter.There was blood in the balloon.There were numerous hypotube kinks.The shaft was buckled throughout the distal shaft.Functional testing was performed by applying pressure with a water filled inflation device and found that water was leaking at the most proximal buckled location.Microscopic examination of the shaft revealed a hole in the outer shaft near the distal end of the core wire which indicates at some point the shaft buckled causing the core wire to puncture through the outer shaft.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 10jan2020.It was reported that a kink occurred.The 99% stenosed, 16mmx2.75mm target lesion was located in the severely tortuous and calcified left main coronary artery.A 2.75mm x 12mm quantum maverick balloon catheter was advanced for dilatation.However, the balloon delivery shaft was kinked.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed a hole in the outer shaft.
|
| |
|