PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Catalog Number 99400-000372
Medical Device Problem Code Failure to Deliver Shock/Stimulation (1133)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/04/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Physio-control contacted the customer to request additional information on the patient.The customer informed physio-control that no further patient information is available.Patient fields in which information is not provided were intentionally left blank.Physio-control evaluated the customer¿s device but was unable to duplicate the reported issue.After observing proper device operation through functional and performance testing the device was returned to the customer for use.The cause of the reported issue could not be conclusively determined.
 
Event or Problem Description
The customer contacted physio-control to report that their device failed to shock during a patient event.The customer advised that the device was able to shock the patient successfully on the first shock, on the second and third shock attempt, the shock was not delivered.On the fourth and fifth shock, the device delivered the energy.This issue is patient related; however there was no adverse patient outcome reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key9643971
Report Number0003015876-2020-00131
Device Sequence Number1965066
Product Code MKJ
UDI-Device Identifier00721902148460
UDI-Public00721902148460
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K102972
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 01/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number99400-000372
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/07/2020
Initial Report FDA Received Date01/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-