| Catalog Number |
99400-000372 |
| Medical Device Problem Code |
Failure to Deliver Shock/Stimulation (1133)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Physio-control contacted the customer to request additional information on the patient.The customer informed physio-control that no further patient information is available.Patient fields in which information is not provided were intentionally left blank.Physio-control evaluated the customer¿s device but was unable to duplicate the reported issue.After observing proper device operation through functional and performance testing the device was returned to the customer for use.The cause of the reported issue could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device failed to shock during a patient event.The customer advised that the device was able to shock the patient successfully on the first shock, on the second and third shock attempt, the shock was not delivered.On the fourth and fifth shock, the device delivered the energy.This issue is patient related; however there was no adverse patient outcome reported.
|
| |
|