| Model Number |
CSE-P-225 |
| Medical Device Problem Codes |
Crack (1135); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Death (1802)
|
| Date of Event |
12/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer went onsite to gather data on the procedure and discovered the leak occurred on cycle 4, with the total volume reinfused of 625ml during the procedure.The bowl was returned to haemonetics for evaluation.Evaluation of the returned bowl confirmed that a leak did occur in the inner core where the bowl base joins with the outer part of the inner core.This could be the possible cause of the leak at the dynamic seal however, it was not possible to confirm due to the condition of the returned sample.There is nothing at this time that would indicate that the death was related to the machine used.It was mentioned by the customer used another cell saver and a new set was used, to continue the procedure, but no further information was provided.
|
| |
|
Event or Problem Description
|
|
On 16 december 2019: haemonetics was informed by the customer of a patient death the next day after surgery.During a surgery with the cell saver elite processing set emitted an alarm for "liquid in the centrifuge".It was reported there was a loss of 250ml in the bowl and from the bowl reservoir.The leak occurred on cycle 4, with a total volume reinfused of 625ml during the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The following sections were updated to reflect the results of the device evaluation: b1: product problem checked.The investigation determined the device did malfunction.H6: codes for device, results and conclusion were updated to reflect the investigation.The leak in the bowl could contribute to the error liquid in the centrifuge received by the customer during the procedure.The most likely cause of the leak is a crack in the inner core.However, cause of the device malfunction contributing to the adverse event could not be established based on the available information provided from the customer.
|
| |
|
Search Alerts/Recalls
|