• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012447-20
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 12/05/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that before use and during device preparation, the protective sheath could not be removed from the nc trek.There was no patient involvement and no clinically significant delay in procedure.A non-abbott device was used to continue the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Visual, dimensional and functional inspections were performed on the returned device.The reported difficulty removing the protective sheath was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported difficulty removing the protective sheath.However, factors that may contribute to difficulty removing the protective sheath may include, but not limited to, manufacturing, normal variation within the manufacturing process or protective sheath removal technique.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9645197
Report Number2024168-2020-01015
Device Sequence Number2139453
Product Code LOX
UDI-Device Identifier08717648151774
UDI-Public08717648151774
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2021
Device Model Number1012447-20
Device Catalogue Number1012447-20
Device Lot Number81217G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2020
Initial Date Received by Manufacturer 01/09/2020
Supplement Date Received by Manufacturer03/31/2020
Initial Report FDA Received Date01/29/2020
Supplement Report FDA Received Date04/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-