• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. BOOT ASSEMBLY; STEREOTAXIC DEVICE, ROBOTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. BOOT ASSEMBLY; STEREOTAXIC DEVICE, ROBOTICS Back to Search Results
Model Number 210080
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/07/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
 
Event or Problem Description
During joint assessment of surgery, surgeon noticed that carbon fibers on the boot appeared to be frayed.Frayed area is noticeable at the top portion of the boot near the calf.No adverse consequences.Please replace boot as soon as possible.Case type: tka.
 
Additional Manufacturer Narrative
Reported issue during joint assessment of surgery, surgeon noticed that carbon fibers on the boot appeared to be frayed.Frayed area is noticeable at the top portion of the boot near the calf.No adverse consequences.Please replace boot as soon as possible.Product inspection: visual inspection confirms the carbon fibers have splintered on the boot assembly.See attached image.Product history review: review of the product history records indicate that under lot no 201843071007: 1.(b)(4) devices were manufactured and accepted into final stock on 07/18/2019.No non-conformances were identified during inspection.2.(b)(4) devices were manufactured and accepted into final stock on 09/10/2019.No non-conformances were identified during inspection.3.(b)(4) devices were manufactured and accepted into final stock on 10/30/2019.No non-conformances were identified during inspection.4.(b)(4) devices were manufactured and accepted into final stock on 11/20/2019.No non-conformances were identified during inspection.Complaint history review: a review of complaints in catsweb and trackwise related to p/n 210080, lot number 201843071007, shows 03 additional complaints related to the failure in this investigation.Pr id (b)(4).Conclusion: the failure was confirmed as alleged via visual inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.Corrective action/preventive action: a review of stryker¿s nc/capa database indicated there has been no nc or capa associated with the product and failure mode reported in this event.
 
Event or Problem Description
During joint assessment of surgery, surgeon noticed that carbon fibers on the boot appeared to be frayed.Frayed area is noticeable at the top portion of the boot near the calf.No adverse consequences.Please replace boot as soon as possible.Case type: tka.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BOOT ASSEMBLY
Common Device Name
STEREOTAXIC DEVICE, ROBOTICS
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key9646676
Report Number3005985723-2020-00058
Device Sequence Number11483988
Product Code OLO
UDI-Device Identifier00848486031879
UDI-Public00848486031879
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number210080
Device Catalogue Number210080
Device Lot Number201843071007
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 01/07/2020
Supplement Date Received by Manufacturer03/12/2020
Initial Report FDA Received Date01/30/2020
Supplement Report FDA Received Date03/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-