| Model Number |
HC445 |
| Medical Device Problem Code |
Misassembled (1398)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
12/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).Age and date of birth unknown / not provided.Patient sex unknown / not provided weight unknown / not provided.First/given name: unknown / not provided.Email address unknown / not provided.
|
| |
|
Event or Problem Description
|
|
Doctor indicated that it was impossible to screw the healing collar.Procedure was completed with another abutment.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmerbiomet complaint number (b)(4).One healing collar (hc445) was returned for investigation.Visual evaluation of the as returned product identified damaged threads and signs of wear.Functional testing was performed using an in-house implant.The device could not seat on the implant because the threads were damaged.Pre-existing patient factors, implant age, tooth location are not relevant to the reported event.Doctor was attempting to place the device on an unknown tooth location when the incident occurred.X-ray or picture images were not provided and no other information was provided.Dhr review for the lot (2019030801) had revealed no deviations nor non-conformances which could have caused or contributed to the reported event.All products were conforming at the time they left zimmer biomet.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the lot (2019030801) for similar event and no other complaint was identified.Therefore, based on the available information and functional testing, device malfunction has occurred and the reported event was confirmed.
|
| |
|
Search Alerts/Recalls
|