| Model Number |
140.640 |
| Medical Device Problem Code |
Loss of Osseointegration (2408)
|
| Health Effect - Clinical Codes |
Fistula (1862); Pain (1994); Tissue Damage (2104)
|
| Date of Event |
12/16/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The information provided in this report was based on information given by the dentist in neodent¿s products warranty form that was recorded in the system and is used by both the manufacturer and the subsidiary.The original complaint and the product were received by the subsidiary and did not arrive in the manufacturer yet.When this happens, a new evaluation of the complaint will be carried out and, if necessary, a new follow-up report will be send.
|
| |
|
Event or Problem Description
|
|
(b)(4)¿ the dentist reported that 3 years and 6 months after the dental implant was installed in intraoral region 23#, its loss of osseointegration was observed.Moreover, 60n.Cm of primary stability was achieved, the patient presented bone type iv.
|
| |
|
Search Alerts/Recalls
|