| Model Number |
3000 |
| Medical Device Problem Codes |
Computer Software Problem (1112); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Software version: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that there was a low output current message received.It was reported that the patient's output current and magnet current had to be lowered in order for the low output current message not to appear.Internal data of the generator was reviewed and indicated that the generator was programmed off and then programmed back on.A system diagnostic test was not recorded within the available internal data of the generator.Internal testing and data review identified that false low output current messages can occur when m3000 (b)(4) software programs a m103-106 generator after a specific programming sequence occurs.When m3000 (b)(4) tablets initiate a programming session with a m103-106 generator, where its output current is programmed off (0 ma), and then the output current is programmed back on to >0 ma, a low output current message would be seen when an in-session interrogation is performed.This would persist upon multiple in-session interrogations and if diagnostics weren't performed during the session, show on the session report.Running system diagnostics or ending and restarting the session will confirm functionality of the device, and resolve the ¿output current low¿ error message.The low output current messages in these cases are false and do not impact generator function.No further relevant information has been received to date.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal reference number: (b)(4).Voluntary medical device correction (remedial action) was initiated by the manufacturer on (b)(4) 2020, and the physician was provided a notification letter with recommended actions.
|
| |
|
Search Alerts/Recalls
|