|
Upon receiving the returned instrument, it was confirmed that the cord had burnt into two pieces near the instrument connection side of the instrument.Based on the lot number on the cord, it can be determined that the cord was purchased in (b)(6) 2017.The ifu states that each olsen reusable device is designed to withstand 20 sterilization cycles when properly cared for and sterilized per the instructions for use.The returned cored had been in use for approximately 2.5 years and had significant signs of wear and tear.The fire that occurred is the result of extensive bending where the cord connects to the bipolar forceps.With significant bending in the same spot the internal wires of the cord with begin to wear.There has been a total of 3,711 sold since 2013 with 4 complaints recorded for similar occurrences (0.1%) in all other cases, the cord has been used for a significant amount of time.Based on this information, this can be seen as the final report.If additional information is obtained that alleges any othe patient involvement or need for corrective actions, a follow up report will be submitted.
|