DEPUY SYNTHES PRODUCTS LLC 1MM DBALL,EX,S; DRILLS, BURRS, TREPHINES AND ACCESSORIES (SIMPLE, POWERED)
|
Back to Search Results |
|
| Catalog Number |
S-1D-G1 |
| Medical Device Problem Code |
Unintended System Motion (1430)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products and therapy dates: drill device, attachment device, burr devices ((b)(6) 2020).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an unspecified surgical procedure it was observed that the cutter device was wobbling at the distal end.It was reported that the same drill and same short attachment was used with other burr devices, including an identical spare device, without any wobbling.It was reported that there were any delays in the procedure due to the event.There was patient involvement reported.There were no reports of any injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D10: additional narrative: the actual device was returned for evaluation.Quality engineering evaluated the device and it was determined that the device passed all manufacturing specifications and no failure was identified.Therefore, the reported condition was not confirmed and an assignable root cause was not determined.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|