| Model Number |
DTMA1Q1 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 459888, lead, implanted: (b)(6) 2018.Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory showed the battery is approaching the indicator signifying the time for device replacement.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that there was concerns of early battery depletion on the cardiac resynchronization therapy defibrillator (crt-d).Follow up concluded that no intervention was done.The device remains in use.No patient complications have been reported as a result of this event.
|
| |
|