The device returned to numed for evaluation.A longitudinal tear is confirmed.The balloon was examined at 10x magnification.Nothing was observed that would have caused the balloon burst.This device was being used off-label for an unapproved use.This device is only approved for percutaneous transluminal valvuloplasty of the pulmonary valve.It was being used for iliac stenosis.A comparative catheter was tested for rated burst pressure.It was the same catalog number / size as the complaint catheter, but a different lot number.The balloon was immersed in a body temperature water bath and inflated in 0.5 atm increments until failure.The balloon tore in a longitudinal manner at 3 atm, which is well above the labeled rated burst pressure of 2.0 atm, and nominal pressure of 1.5 atm.As per the instructions for use, an inflation device with pressure gauge should be used on the device to monitor pressure.In this case, an inflation device with pressure gauge was used, however it was unknown as to what pressure the catheter was taken to and what pressure it burst at.A review of the device history record was performed and no issues were noted.There are no other complaints associated with this lot number.A review of the balloon material used shows there are no other complaints associated with the balloon material lot.
|