• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARGON MEDICAL DEVICES INC. OPTION IVC FILTER; RETRIEVABLE INFERIOR VENA CAVA FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARGON MEDICAL DEVICES INC. OPTION IVC FILTER; RETRIEVABLE INFERIOR VENA CAVA FILTER Back to Search Results
Catalog Number 352506070E
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/22/2019
Event Type  malfunction  
Manufacturer Narrative
Sample device was returned to manufacturer on 1/27/2020.Device evaluation is in progress.A follow-up report with additional information will be provided by 2/28/2020.
 
Event Description
Filter did not deploy.
 
Manufacturer Narrative
A review of the sample device was performed.Only the delivery sheath and pusher wire was returned, not the filter.A minor kink was found on the distal and proximal side of the pusher but could be pushed through the delivery sheath without issue.Without all parts of the device system, the complaint could not be duplicated or confirmed.Since the problem could not be determined as a design or manufacturing issue, a corrective action will not be taken at this time.
 
Manufacturer Narrative
The sample device was returned to argon from sterilization on 2/11/2020.Investigation is in progress.A follow-up report will be provided by 3/13/2020.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTION IVC FILTER
Type of Device
RETRIEVABLE INFERIOR VENA CAVA FILTER
Manufacturer (Section D)
ARGON MEDICAL DEVICES INC.
1445 flat creek road
athens TX 75751
MDR Report Key9649932
MDR Text Key182302716
Report Number1625425-2020-00100
Device Sequence Number1
Product Code DTK
Combination Product (y/n)Y
PMA/PMN Number
K133243
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup,Followup
Report Date 01/01/2005,01/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number352506070E
Device Lot Number11274557
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received01/14/2005
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-