• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA BRK¿ XS TRANSSEPTAL NEEDLE, 98 CM LENGTH; TROCAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA BRK¿ XS TRANSSEPTAL NEEDLE, 98 CM LENGTH; TROCAR Back to Search Results
Model Number G407211
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Cardiac Perforation (2513)
Date of Event 01/20/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported effusion remains unknown.Per the ifu, cardiac perforation is a known risk during the use of this device.
 
Event or Problem Description
During transseptal puncture, a pericardial effusion occurred.Contrast medium was used to assist during transseptal and it was noted that the needle was on the septum with tenting, but introduction of the guidewire revealed that it was not in the left atrium.The patient began to feel unwell a few seconds after puncture and was treated with medication before a second successful puncture.After a worsening of the patient's hemodynamic state, an ultrasound revealed a pericardial effusion for which a pericardiocentesis was performed to stabilize the patient.There were no performance issues with any abbott device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK¿ XS TRANSSEPTAL NEEDLE, 98 CM LENGTH
Common Device Name
TROCAR
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
MDR Report Key9652799
Report Number3008452825-2020-00064
Device Sequence Number1742779
Product Code DRC
UDI-Device Identifier05414734205153
UDI-Public05414734205153
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K122587
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Model NumberG407211
Device Catalogue NumberG407211
Device Lot Number7222205
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2020
Initial Report FDA Received Date01/31/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AGILIS¿ NXT STEERABLE INTRODUCER
Outcome Attributed to Adverse Event Required Intervention;
-
-