Model Number 9554
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/08/2019
Event Type  malfunction  
Manufacturer Narrative
Device is a combination product.A promus premier ous mr 38 x 2.75 mm stent delivery system was returned for analysis.A visual examination of the stent found no issues.There were no signs of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal marker bands.The crimped stent outer diameter was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple kinks and a break 22.2cm distal to the distal end of the strain relief.A visual examination of the outer and inner lumen and mid-shaft section found no issues.No other issues were identified during the product analysis.
 
Event Description
Reportable based on device analysis completed on 10jan2020.It was reported that crossing difficulties were encountered.The target lesion was located in the moderately tortuous left anterior descending artery.A 38 x 2.75 promus premier drug-eluting stent was advanced but failed to cross the lesion.The procedure was completed with no patient complications and the patients status was stable.However, returned device analysis revealed hypotube break.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9653984
MDR Text Key178647795
Report Number2134265-2020-00796
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 01/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9554
Device Catalogue Number9554
Device Lot Number0023064076
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/10/2020
Initial Date FDA Received01/31/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/30/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age83 YR
-
-