MAUDE Adverse Event Report: AD-TECH MEDICAL INSTRUMENT CORP. SD DEPTH ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AD-TECH MEDICAL INSTRUMENT CORP. SD DEPTH ELECTRODE Back to Search Results
Model Number SD08R-SP05X-000
Medical Device Problem Code Inaccurate Information (4051)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/15/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Mdr 2183456-2019-00021 has been filed past the 30-day timeframe.Through the 2019 fda inspection, it was identified that while ad-tech was evaluating reportability for actual events in which death or serious injury had occurred, ad-tech had failed to submit events in which a malfunction was reported and could result in a serious injury if the event were to recur.As part of the investigation, a two (2) year retrospective review of complaints was performed to determine which complaints required submission of an mdr.
 
Event or Problem Description
On april 15, 2019, an ad-tech clinical specialist reported that during a case they attended the surgeon encountered an issue after implanting four ad-tech depth electrodes.Upon initial device testing subsequent to implantation it was identified that two of the devices that were labeled as 8 contact depth electrodes were in fact 6 contact depth electrodes.There was no impact to the patient due to this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SD DEPTH ELECTRODE
Common Device Name
DEPTH ELECTRODE
Manufacturer (Section D)
AD-TECH MEDICAL INSTRUMENT CORP.
400 west oakview parkway
oak creek WI 53154
MDR Report Key9654450
Report Number2183456-2019-00021
Device Sequence Number1709139
Product Code GZL
UDI-Device Identifier00841823107695
UDI-Public(01)00841823107695(17)210101(10)0127219208140699
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K163355
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/01/2021
Device Model NumberSD08R-SP05X-000
Device Catalogue NumberSD08R-SP05X-000
Device Lot Number208140699
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/15/2019
Initial Report FDA Received Date01/31/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-