• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER AC ADAPTER; VENTRICULAR (ASSIST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER AC ADAPTER; VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number 1430US
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/13/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the controller ac adapter did not provide power.The controller ac adapter was exchanged.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
### a supplemental report is being submitted for device analysis.Product event summary: the controller ac adapter was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Visual inspection revealed a missing clamp assembly screw from the ac inlet clamp assembly.This is an additional finding not related to the reported event.The most likely root cause of the missing clamp assembly screw can be attributed to the handling of the device.Functional testing revealed that the controller ac adapter was unable to provide power to the test controller.Supplemental testing revealed that the pulse width modulation (pwm) controller was damaged.After replacing the defective component, the adapter performed as expected.As a result, the reported event was confirmed.Based on the available information, the most likely root cause of the reported event can be attributed to a faulty component.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.### medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER AC ADAPTER
Common Device Name
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key9655944
Report Number3007042319-2020-00347
Device Sequence Number19302243
Product Code DSQ
UDI-Device Identifier00888707002738
UDI-Public00888707002738
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/30/2021
Device Model Number1430US
Device Catalogue Number1430US
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2020
Initial Date Received by Manufacturer 01/13/2020
Supplement Date Received by Manufacturer05/12/2020
Initial Report FDA Received Date01/31/2020
Supplement Report FDA Received Date05/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1103 VAD
Patient Age41 YR
-
-