| Model Number |
MMT-751NAH |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
08/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Insulin pump passed the rewind, basic occlusion, prime/a33 and displacement test.Pump received with stuck motor error alarm loop during bolus delivery.The motor was tested outside of the device and passed.The motor may have had and intermittent failure that was not detected during our testing.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had motor error alarm.Customer reported that the drive support cap appears to be normal.Insulin pump not exposed to high magnetic field.Customer was able to successfully clear the alarm.Customer was able to complete insulin pump rewind.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|